Pressure sensors are used in dynamic applications, such as in the aerospace industry. However, these pressure transducers are mostly calibrated using static methods. SPEKTRA LABS offers you various solutions for the calibration of sensors used for dynamic pressure measurement. We can dynamically excite and reliably calibrate a wide range of pressure sensors.
Your pressure transducers should be calibrated regularly to keep the quality of your measurements at a high level.
Detailed information available for download:
Flexible Accreditation (in German)

Piezoresistive pressure sensors
- Dynamic calibration and testing using air pressure
Frequency range: 10 Hz…1250 Hz
Amplitude: 0.2 kPa…12 kPa (0.0002 bar…0.12 bar) - Dynamic calibration using hydraulic pressure pulses
Frequency range: up to 10 kHz
Amplitude: 15 bar - Overload / load tests
Amplitude: 20 bar…500 bar